The specified lifetime of the thermal block cycler is 105,000 cycles (corresponding to approx. 4 runs with 43 cycles on 200 days over 3 years).
Temperature Homogeneity
The thermal block cycler of the LightCycler® 480 Instrument has three different temperature sensors: one sensor is located in the center of the plate mount and is used for absolute temperature control; the other two sensors for checking temperature homogeneity are located on top of the left and right section of the block.
No Recalibration of the LightCycler® 480 Instrument after exchange of the thermal block cycler unit
Each thermal block cycler unit has its own EEPROM*, carrying block-specific information, such as calibration data. During instrument start-up, information within the EEPROM is read by the instrument controller and transferred to the LightCycler® 480 Software. (*EEPROM = Electrically Erasable Programmable Read-Only Memory)
General Cleaning of the Thermal Block Cycler
Regular cleaning of the LightCycler® 480 Instruments and accessories is not required. If necessary, clean the housing of the LightCycler® 480 Instrument, the thermal block cycler, and the block cycler cover with a mild commercial detergent. Use 70% ethanol for disinfecting the instrument housing, the thermal block cycler, and the block cycler cover.
Cleaning of the LightCycler® 480 block cycler unit: Pipette into all wells 20 μl (384-well block), respectively 125 μl (96 well block), 70% ethanol or isopropanol. After 15 minutes, pipette up and down several times. Remove the liquid and let the block cycler unit dry before use. Take care not to degrade the block coating during cleaning. ",
"Country": "XG",
"Code": "Background Information - Help Corner",
"Name": "Background Information - Help Corner"
},
{
"Language": "en",
"Value": "
Thermal Block Cycler Unit for 96-well PCR plates
Corresponding Block Lid for 96-well PCR plates
Storage Box for both accessories
",
"Country": "XG",
"Code": "Content",
"Name": "Content"
},
{
"Language": "en",
"Value": "The LightCycler® 480 Block Kit 96 Silver is an interchangeable accessory for the LightCycler® 480 Instrument enabling users to perform 96-well format applications (sample volume: 10 to 100 μl).",
"Country": "XG",
"Code": "Applications",
"Name": "Applications"
},
{
"Language": "en",
"Value": "The LightCycler® 480 Instrument is a versatile and customizable PCR device, allowing the use of different, easily interchangeable thermal block cycler units (96-well and 384-well format). The LightCycler® 480 thermal block cycler unit is provided in a storage box which protects the accessory from damage that could occur during transport. The storage box contains the thermal block cycler unit assembled together with the block lid, inside the block cycler loading device.",
"Country": "XG",
"Code": "Product Description",
"Name": "Product Description"
}
]
}
}
]
}
LightCycler® 480 Block Kit 96 Silver
Easily interchangeable 96-well thermal block cycler unit for the LightCycler® 480 Instrument.